Skip to main content
. 2018 Dec;10(6):433–440.

Table 3.

Comparison of CSB and KOH as reference method according to CFW stain as Gold.St method for diagnosis of fungal elements in 30 minutes incubation (N= 189)

Test Positive Cases N (%) Sensitivity (%) Specificity (%) PPVa (%) NPVb (%) Accuracy (%)
KOH 68 (36) 66 98 97 98 80
CSB 104 (55) 97 100 100 100 98
a

Positive predictive values (PPV)

b

Negative predictive values (NPV)