Table 1.

Results of the detection of PSMA in mouse serum by our hydrophilic MIP-based Love wave sensor.

Sample PSMA Added (ng mL−1) PSMA Detected (ng mL−1) Recovery (%)
1 1 0.96 ± 0.11 96
2 10 9.54 ± 1.0 95.4
3 20 18.75 ± 1.9 93.8
4 30 31.09 ± 3.0 103.6
5 50 51.96 ± 5.4 103.9
6 60 57.32 ± 6.9 95.5
7 80 82.43 ± 7.3 103
8 100 103.51 ± 9.8 103.5