Skip to main content
. 2018 Nov 19;56(2):210–218. doi: 10.1177/0004563218812986

Table 1.

LOD, LOQ, RSD and urine FA concentrations detected by using FA probe and Fluo-HPLC methods within a day at different time points, respectively.

Test items 0 min 30 min 1 h 2 h Mean ± S.D. Recovery (%) RSD (%) LOD (μM) LQD (μM)
FA probe-urine FA 94.52 96.85 92.17 91.03 93.64 ± 6.26 97.3 6.68 1.27 2.48
HPLC-urine FA 52.49 53.33 53.58 54.06 53.36 ± 3.85 98.1 7.22 1.52 2.91

Note: Data in means (n = 10). No significant difference among the values detected by these two methods within a day (P > 0.05), respectively.

FA: formaldehyde; LOD: limits of detection; LOQ: limits of quantization; RSD: relative standard deviations; HPLC: high-performance liquid chromatography.