Table 5.
The sensitivity and specificity of the VDN LAMP when compared to the arpV2/aprB2 rtPCR of moist (M) or dry (D) samples collected from 309 individual sheep.
SM | Sheep properties (n) | Sheep (n) | rtPCR aprV2 + (n) | VDN LAMP + (n) | Se (%) | Sp (%) | Cohen's kappa (agreement) |
---|---|---|---|---|---|---|---|
M | 7 | 127 | 68 | 59a | 83.82 | 96.61 | 0.796 good |
D | 12 | 182 | 92 | 38b | 39.13 | 97.78 | 0.367 fair |
2 samples VDN LAMP false positive.
2 samples VDN LAMP false positive.