. 2019 Mar 14;9:4494. doi: 10.1038/s41598-019-40960-5

Table 2.

Detection limit of the ZIKV RT-LAMP assaya.

ZIKV Concentration (PFU) No. of Replicates No. of positive results Hit rate in %
105 10 10 100
104 10 10 100
103 10 10 100
102 10 10 100
101 10 10 100
100 10 10 100
10−1 10 10 100
10−2 10 10 100
10−3 10 9 90
10−4 10 7 70
10−5 10 6 60
10−6 10 0 0
10−7 10 0 0

aProbit regression analysis was calculated using MedCalc software (version 18.11), giving a C95 value (concentration detectable 95% of the time) of −2,98 log10 PFU of ZIKV. This indicates that the limit of detection is about −3 log10 (1/1000) PFU/reaction and that samples containing that concentration would be detected 95% of the time.