Skip to main content
. 2019 Mar 15;11:51. doi: 10.1186/s13148-019-0642-0

Table 3.

Precision evaluation results of the meSDC2 LTE-qMSP test kit

Target HCT116 genomic DNA concentration (pg) Mean CT (%CV)
Reproducibility Repeatability Lot-to-lot variation
Site 1 Site 2 Lot 1 Lot 2
SDC2 100 26.19 (2.71) 25.75 (2.10) 26.1 (2.68) 26.19 (2.71) 25.75 (2.10)
500 23.52 (2.21) 23.15 (1.12) 23.44 (2.13) 23.52 (2.21) 23.15 (1.12)