TABLE 3.

Projected impact of observed proportional bias on interpretation

Resistance Measurements 4-year-old male (105cm) 8.5-year-old male (129cm)
Value measured on IOS kPa·L−1·s−1 Projected value measured on tremoFlo# kPa·L−1·s−1 IOS value z-score tremoFlo value z-score Difference z-score+ IOS value z-score tremoFlo value z-score Difference z-score+
R5 0.75 0.73 −1.35 −1.44 0.09 0.20 0.10 0.09
1 1.01 −0.27 −0.22 −0.05 1.28 1.33 −0.05
1.25 1.29 0.82 1.01 −0.19 2.37 2.56 −0.19
1.5 1.58 1.90 2.24 −0.33 3.45 3.79 −0.33
R10 0.75 0.85 −0.71 −0.28 −0.43 0.68 1.11 −0.43
1 1.19 0.39 1.24 −0.85 1.79 2.64 −0.85
1.25 1.54 1.50 2.77 −1.27 2.90 4.17 −1.27
1.5 1.88 2.61 4.30 −1.69 4.00 5.69 −1.69

IOS: MasterScreen impulse oscillometry system. #: based on the observed average resistance- and oscillatory frequency-specific proportional bias (note: we conservatively used the bias observed in 38 children, without removing the outlier identified for R5); : z-score values calculated using IOS-specific paediatric normative values [9]; +: z-score difference between those derived from the IOS versus tremoFlo, with negative values indicating a lower z-score value on the IOS than the tremoFlo.