Continuous Glucose Monitoring Performance Across Days of Sensor Wear

Clinic session day Matched pairs (n) %15/15 (%) %20/20 (%) %30/30 (%) MARD (%)
Day 1 719 81.5 92.2 99.0 9.3
Day 2 664 84.0 92.3 100.0 8.4
Days 4–5 880 85.8 95.5 99.2 9.4
Day 7 588 82.1 97.3 99.8 8.7
Day 10 681 82.4 92.5 99.1 9.0