Table 6.
Reliability index definitions (false positive, false negative, and fixation loss)
| Reliability index | Humphrey Field Analyzer SITA Standard | imo® AIZE |
|---|---|---|
| False positive | Conversion from reaction time (response within 0.2 seconds) | Ratio of responding by presenting a 9 dB dark target to the determined threshold Binocular random conditions |
| False negative | Ratio of not responding by presenting a 9 dB bright target to the determined threshold Included observation point unknown Monocular | Ratio of not responding by presenting a 9 dB bright target to the determined threshold All observation points known except for outermost circumference Binocular random conditions |
| Fixation loss | Percentage in response to stimulus to blind spots | Percentage in response to stimulus to blind spots |