Skip to main content
. 2019 Mar 18;9:4800. doi: 10.1038/s41598-019-41344-5

Table 5.

Diffusion-weighted MR scan protocol. (*) Philips Ingenia and GE Discovery.

Lung Cancer (site A, B, E, G) Colorectal liver metastases (1.5 T) (site A, B, C, E) Colorectal liver metastases (3 T) (site C, F*) Ovarian cancer (site E)
Sequence ss-EPI ss-EPI ss-EPI ss-EPI
TR (ms) ≥8000 ≥8000 5000 ≥8000
TE (ms) minimum minimum minimum minimum
NSA 4 4 2–4 4
FOV (mm2) 380 × 273 380 × 380 380 × 273 332 × 380
Matrix 128 × 112 128 × 128 128 × 128 128 × 112
Bandwidth (Hz/px) 1400–1800 1400–1800 1500–2650 1400–1800
Slice thickness (mm) 5 5 5–6 6
Slice gap (mm) 0 0 0 0
Pixel size (mm2) 3 × 3 1.5 × 1.5 1.5 × 1.5 1.5 × 1.5
b-values 100, 500, 800 100, 500, 900 150, 400, 800 100, 500, 900
Fat saturation yes yes yes yes
Parallel imaging yes yes yes yes