. 2018 Dec 8;16:100174. doi: 10.1016/j.jcte.2018.12.003

Table 3.

Statistical analysis of agreement in gestational diabetes diagnosis between the glucometers against the hexokinase method laboratory reference.

All Glucometers (95% CI) Contour Plus (Bayer) (95% CI) Freestyle Optium Neo (Abbott) (95% CI) Accu-Chek Active(Roche) (95% CI) Glucocheck Classic (TaiDoc) (95% CI) One touch Select Plus Flex (Lifescan) (95% CI)
Number of participants 592 143 143 56 105 145
True Positive (n) 27 7 8 1 9 2
False Positive (n) 48 5 6 1 26 10
True Negative (n) 514 131 129 53 70 131
False Negative (n) 3 0 0 1 0 2
Sensitivity (%) 90 (74; 98) 100 (59; 100) 100 (63; 100) 50 (1; 99) 100 (66; 100) 50 (7; 93)
Specificity (%) 91 (89; 94) 96 (92; 99) 96 (91; 99) 98 (90; 100) 73 (63; 81) 93 (87; 97)
Positive Predictive Value (%) 36 (25; 48) 58 (28; 85) 57 (29; 82) 50 (1; 99) 26 (12; 43) 17 (2; 48)
Negative Predictive Value (%) 99 (98; 100) 100 (97; 100) 100 (97; 100) 98 (90; 100) 100 (95; 100) 98 (95; 100)
Positive Likelihood Ratio 10.54 27.20 22.50 27.00 3.69 7.05
Negative Likelihood ratios 0.11 0.00 0.00 0.51 0.00 0.54

CI = confidence interval.