Table 2. Agreement of results for detection of metastases in slides scanned with the miniature microscope scanner and the reference slide scanner.
Device | Ground Truth | Momic (Expert 1) | Reference scanner (Expert 1) | Momic (Expert 2) | Reference scanner (Expert 2) |
---|---|---|---|---|---|
Ground Truth | 1 | ||||
Momic (Expert 1) | k = 0.95 (CI95%: 0.90–1.00) | 1 | |||
Reference scanner (Expert 1) | k = 0.95, (CI95%: 0.90–1.00) | k = 0.94 (CI95%: 0.88–1.00) | 1 | ||
Momic (Expert 2) | k = 0.87, (CI95%: 0.79–0.96) | k = 0.92 (CI95%: 0.84–0.99) | k = 0.86 (CI95%: 0.77–0.95) | 1 | |
Reference scanner (Expert 2) |
k = 0.94, (CI95%: 0.88–1.00) | k = 0.95 (CI95%: 0.90–1.00) | k = 0.92 (CI95%: 0.85–0.99) | k = 0.92 (CI95%: 0.85–0.99) | 1 |
Comparison of analysis of slides. Agreement calculated using kappa statistics with associated confidence intervals (95%); p < 0.001 for all values shown.