Table 2. Agreement of results for detection of metastases in slides scanned with the miniature microscope scanner and the reference slide scanner.

Device Ground Truth Momic (Expert 1) Reference scanner (Expert 1) Momic (Expert 2) Reference scanner (Expert 2)
Ground Truth 1
Momic (Expert 1) k = 0.95 (CI95%: 0.90–1.00) 1
Reference scanner (Expert 1) k = 0.95, (CI95%: 0.90–1.00) k = 0.94 (CI95%: 0.88–1.00) 1
Momic (Expert 2) k = 0.87, (CI95%: 0.79–0.96) k = 0.92 (CI95%: 0.84–0.99) k = 0.86 (CI95%: 0.77–0.95) 1
Reference scanner
(Expert 2)
k = 0.94, (CI95%: 0.88–1.00) k = 0.95 (CI95%: 0.90–1.00) k = 0.92 (CI95%: 0.85–0.99) k = 0.92 (CI95%: 0.85–0.99) 1

Comparison of analysis of slides. Agreement calculated using kappa statistics with associated confidence intervals (95%); p < 0.001 for all values shown.