Table 3. Patient cases diagnosed incorrectly in analysis of WSIs from both scanners, compared to light microscopy.

Case number Staining type of sample Light microscopy diagnosis
(Ground truth)
MoMic WSI (Expert 1) Reference slide scanner WSI (Expert 1) MoMic WSI (Expert 2) Reference slide scanner WSI (Expert 2)
1 Toluidine blue Micrometastasis No metastasis (FN) Metastasis No metastasis (FN) No metastasis
IHC Micrometastasis No metastasis (FN) Metastasis Metastasis Metastasis
2 Toluidine blue Micrometastasis No metastasis (FN) No metastasis (FN) No metastasis (FN) No metastasis (FN)
IHC Not available Not available Not available Not available Not available
3 Toluidine blue Macrometastasis Metastasis Metastasis No metastasis (FN) Metastasis
IHC Macrometastasis Metastasis Metastasis No metastasis (FN) Metastasis
4 Toluidine blue No metastasis Not evaluable No metastasis No metastasis No metastasis
IHC No metastasis Not evaluable No metastasis No metastasis No metastasis

Table showing discrepant patient cases, compared to light microscopy. Included also one case with both miniature microscope scanner WSIs classified as not evaluable by one expert (number 4).