Skip to main content
. 2019 Feb 27;57(3):e01714-18. doi: 10.1128/JCM.01714-18

TABLE 2.

Results of low-copy plasma standards used for calculating the 95% LoD of iSCA v2.0

HIV-1 RNA plasma standard (cps/ml) No. of replicates tested No. of replicates detected Replicates detected (%) Mean HIV-1 RNA cps expected Mean HIV-1 RNA cps detected Mean HIV-1 RNA (cps/ml)
4 10 10 100 20 23.7 4.7
3 10 10 100 15 19.0 3.8
2 10 10 100 10 11.6 2.3
1 20 19 95 5 6.4 1.3
0.3 20 12 60 1.5 3.8 1.2
0.1 20 4 20 0.5 0.52 0.5