TABLE 4.

Sensitivity, specificity, positive predictive value, and negative predictive value of FP-VLP-MAC-ELISA and NS1-MAC-ELISA for the validation serum panel based on the developed algorithma

Assay Disease status Disease prevalence Sensitivity Specificity PPV NPV
FP-VLP-MAC-ELISA ZIKV 33.30 (21.7–46.7) 100 (83.2–100) 80 (64.4–91.0) 71.40 (57.4–82.3) 100
DENV 75 (50.9–91.3) 92.50 (79.6–98.4) 83.30 (62.1–93.9) 88.10 (77.5–94.1)
NS1-MAC-ELISA ZIKV 33.30 (21.7–46.7) 80 (56.3–94.3) 95 (83.1–99.4) 88.90 (67.1–96.9) 90.50 (79.8–95.8)
DENV 100 (83.2–100) 92.50 (79.6–98.4) 87.00 (69.2–95.2) 100
a

PPV, positive predictive value; NPV, negative predictive value. Values are percentages, with 95% confidence intervals given in parentheses.