Figure 6.

Figure 6

Clarke error grid analysis on blood glucose concentrations obtained with the tear-glucose device and a clinically approved glucometer. All data were plotted within zones A and B, showing the reliability of the blood glucose concentrations estimated with the tear-glucose device29,30. The y-axis represents the blood glucose concentration estimated from tear glucose concentration measured with the tear-glucose device, and the x-axis represents the blood glucose concentration measured with a reference method using a commercially available glucometer. Region A indicates that the data are within 20% error from the reference method. Region B indicates that the values are over the 20% error mark, but do not imply an inappropriate treatment. Region C indicates that the data will lead to an unnecessary treatment. Region D indicates that the data will fail to diagnose hyperglycemia or hypoglycemia. Region E indicates that the data will lead to a treatment for hypoglycemia to a patient with hyperglycemia, or vice versa.