Table 9.

Various artefacts in EEG signal recording

Category Artefact /Source(Cause)/Frequency/Amplitude Morphology Artefact Prevention
Physiological
Artefacts
Cardiac/Heart/1Hz/1-10mV
/Epilepsy
Selection of proper montage
Monitoring during recording
Offline visual inspection
Low pass filter (LPF)
Data Rejection
EOG/Eye/0.5-3 Hz/100mV
/Tumour, delta wave
Artefact-free recording protocol
Online monitoring
Offline visual inspection
Low pass filter (LPF)
Using various
Transform (ICA, PCA, EOG subtraction)
Muscle Artefact/Muscle
/ 100Hz/low /Beta frequency
Artefact-free recording protocol
Online Monitoring
Offline visual inspection
High pass filter(HPF)
Data rejection
Physical movement artefact
/Physical movement
/Very low/ very high
/Morphology different
from actual EEG
Artefact-free recording protocol
Online monitoring
Offline visual inspection
Data rejection
External
Artefacts
Transmission line
/Transmission line
50–60 Hz/low/
Morphology different
from actual EEG
Notch filter
DC power supply
Phone artefacts
/Mobile and landline phone
/high/high/different
Artefact-free recording protocol
Electrode artefact
/Electrode and sweating
/very low/high
Artefact-free recording protocol
LPF
Impedance artefact
/Electrode with impedance>5KΩ
/-/approximately 100 μV/different