Table 2.

Precision of the Diasys, Evermed, Hengxiao, Diazyme, and Zybio assays

Manufacturer Pool‐1 Pool‐2
Mean (U/L) (±SD) Within‐laboratory CV% Interassay CV% Mean (U/L) (±SD) Repeatability CV% Within‐laboratory CV%
Diasys 512 ± 10.5 1.3% 1.6% 1068 ± 24.5 1.6% 1.9%
Evermed 524 ± 7.7 0.9% 1.6% 1094 ± 21.5 1.6% 2.9%
Hengxiao 151 ± 2.3 1.6% 2.6% 247 ± 5.0 2.6% 1.9%
Zybio 597 ± 10.6 1.1% 1.1% 1045 ± 24.0 1.1% 1.9%
Diazyme 140 ± 2.0 0.8% 1.1% 195 ± 2.1 1.1% 2.0%

Lp‐PLA2 activity is present as the means ± standard deviations using two known Lp‐PLA2 activities for residual fresh serum. Repeatability and within‐laboratory CV% are shown as the percentage of the coefficient of variation.

CV, coefficient variation.