Skip to main content
. 2018 Aug 13;33(1):e22645. doi: 10.1002/jcla.22645

Table 3.

Performances of real‐time PCR assays for detection of nontuberculous mycobacteria depending on culture results

PCR Assay NTM growth NTM no growth Sensitivity %, (95% CI) Specificity %, (95% CI)
PCR + PCR − PCR + PCR −
All AdvanSure 18 2 1 122 90.0 (68.3‐98.8) 99.2 (95.6‐100)
PowerChek 18 2 2 121 90.0 (68.3‐98.8) 98.4 (94.3‐99.8)
Smear + AdvanSure 14 0 0 37 100 (76.8‐100) 100 (90.5‐100)
PowerChek 14 0 0 37 100 (76.8‐100) 100 (90.5‐100)
Smear − AdvanSure 4 2 1 85 66.7 (22.3‐95.7) 98.8 (93.7‐100)
PowerChek 4 2 2 84 66.7 (22.3‐95.7) 97.7 (91.9‐99.7)

CI, confidence interval; NTM, nontuberculous mycobacteria.