Table 2.

Results of preliminary optimization.

Column Mobile phase Elution mode Flow rate Observation Result
Luna C8 Methanol-0.002 M solution of tetrabutylammonium sulfate (20 : 80, v/v) Isocratic 1.0 mL/min Elution force was too weak for sodium benzoate (retention time of sodium benzoate was more than 30 minutes) Rejected

Luna C8 Acetonitrile-0.002 M solution of tetrabutylammonium sulfate (20 : 80, v/v) Isocratic 1.0 mL/min The two peaks of potassium guaiacolsulfonate were not completely resolute; retention time too long for sodium benzoate (more than 20 minutes) Rejected

Luna C8 Methanol-0.002 M solution of tetrabutylammonium sulfate (50 : 50, v/v) Isocratic 1.0 mL/min Unable to distinguish the two components of potassium guaiacolsulfonate Rejected

Luna C8 Methanol-0.005 M solution of tetrabutylammonium hydrogen sulfate (20 : 80, v/v) Isocratic 1.0 mL/min Poor and unstable peak shape for both analytes Rejected

Luna C8 Methanol-0.002 M solution of tetrabutylammonium sulfate (20 : 80, v/v) Gradient (the gradient program was presented in Table 1) 1.0 mL/min Good resolution and peak shape for components of potassium guaiacolsulfonate; good resolution between potassium guaiacolsulfonate and sodium benzoate; acceptable analysis time (15 minutes) Accepted