Skip to main content
. 2019 Mar 7;2019:6143061. doi: 10.1155/2019/6143061

Table 7.

Results of system precision for sodium benzoate.

No. of injection Retention time of sodium benzoate (minutes) Peak area (mAu·s) Asymmetry of peak Number of theoretical plates Resolution
1 12.340 1905008 0.99 27071 14.0
2 12.386 1906997 0.99 27273 14.1
3 12.416 1906631 0.99 27405 14.2
4 12.344 1904799 0.98 27088 14.1
5 12.291 1905328 0.98 26856 14.0
6 12.334 1905038 0.99 27044 14.2
Average 12.352 1905634 0.99 27123 14.1
RSD (%) 0.4 0.1 0.5 0.7 0.6