. 2019 Mar 10;2019:1417967. doi: 10.1155/2019/1417967

Table 3.

Diagnostic sensitivity, specificity, PPV, NPV, and ROC analysis, and kappa values of agreement between test methods.

Axillary temp
(>37.5°C)
Light microscopy SD Bioline™
Sensitivity % 74.3 94.86 95.33
(number/total) 159/214 203/214 204/214
95%CI 67.90-80.01 90.99-97.41 91.57-97.74

Specificity % 53.77 94.34 94.34
(number/total) 57/106 100/106 100/106
95%CI 43.82-63.51 88.09-97.86 88.09-97.89

NPV % 50.89 90.09 90.91
(number/total) 57/112 100/111 100/110
95%CI 43.72-58.03 83.61-94.18 84.50-94.83

PPV % 76.44 97.13 97.14
(number/total) 159/208 203/209 204/210
95%CI 72.26-80.17 93.96-98.66 93.98-98.67

ROC analysis AUC 0.640 0.946 0.948
Standard error 0.032 0.012 0.012
95%CI 0.585-0.693 0.915-0.968 0.918-0.970

Kappa values 0.277 0.882 0.888

Legend. PPV: positive predictive value, NPV: negative predictive value, ROC: receiver operating characteristic curve, AUC: area under the ROC curve, RDT: rapid diagnostic test, and 95CI: 95% confidence interval. Nested PCR considered reference method.