Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2012 Jun 6;901:25–33. doi: 10.1016/j.jchromb.2012.05.036

Table 1.

Accuracy and precision of MX at LLOQ in six patients’ pre-dosed plasmas (n = 5)

Pre-dosed patient plasma Nominal [MX] (ng/ml) Measured [MX] (ng/ml) SD (ng/ml) %CV %RE
#1 1.25 1.26 0.04 3 0.8
#2 1.25 1.23 0.03 2 −2
#3 1.25 1.24 0.02 2 −0.8
#4 1.25 1.29 0.03 2 3
#5 1.25 1.21 0.03 2 −3
#6 1.25 1.23 0.04 3 −2
Meana 1.25 1.24 0.01 0.8 −0.8
a

X = (x1 + x2 + …x6)/6; SD=((sd12+sd22+sd32+sd42+sd52+sd62)/6; %CV =SD/X; %RE = [(measured value − nominal value)/(nominal value)] × 100%