Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2012 Jun 6;901:25–33. doi: 10.1016/j.jchromb.2012.05.036

Table 2.

Matrix effect on MX QC samples in the pooled blank human plasma (n = 5)

MX QC (ng/ml) PAMX in extracted plasma ± SD (x 105)a PAMX in mobile phase ± SD (x 105) MFMX ± SDb PAIS in extracted plasma ± SD (x 105) PAIS in mobile phase ± SD (x 105) MFIS ± SDb IS Normalized MF ± SDb
3.50  0.40 ± 0.02  0.960 ±0.005  0.42 ± 0.02 5.7 ± 0.2 13.4 ± 0.1  0.42 ± 0.02 1.00 ± 0.07
45.0 6.8 ± 0.3 12.5 ± 0.3  0.54 ± 0.03 7.5 ± 0.4 13.4 ± 0.2  0.56 ± 0.03 0.96 ± 0.07
450 52.0 ± 0.5 101 ± 2  0.51 ± 0.01 6.00 ± 0.09 11.9 ± 0.3  0.50 ± 0.01 1.02 ± 0.03
a

PA = mean peak area

b

y(±sy)=a(±sa)b(±sb),sy=y(saa)2+(sbb)2