Table 5.

Accuracy and precision of MX plasma calibrators over five validation batches

MX calibrator (ng/ml) Measured [MX] (ng/ml) SD (ng/ml) %CV %RE
1.25 1.27 0.02 3 2
2.50 2.44 0.06 2 −2
5.00 4.95 0.1 2 −1
12.5 12.2 0.2 2 −2
25.0 24.3 0.5 2 −3
50.0 51.2 0.4 0.8 2
125 125 1 0.8 0
250 253 1 0.4 1
500 513 7 1 3