Table 7.
The interference study of TMZ on the measurement of MX in human plasma (n = 5)
| MX QC (ng/ml) | 3.50 | 45.0 | 450 | |||
| TMZ (20.0 μg/ml) | Yes | No | Yes | No | Yes | No |
| Measured [MX] (ng/ml) | 3.58 | 3.65 | 46.8 | 47.5 | 477 | 475 |
| SD (ng/ml) | 0.05 | 0.03 | 0.9 | 0.7 | 6 | 6 |
| %CV | 1 | 1 | 2 | 1 | 1 | 1 |
| %RE | 2 | 4 | 4 | 6 | 6 | 6 |