Table 9.

Example of incurred patient sample re-analysis.

Patient sampling schedule (h) Measured [MX] (ng/ml) E (ng/ml)a %REb
First analysis Re-analysis after 9 months
Pre-dose no peak no peak not applicable not applicable
0.250 5.36 5.69 0.33 6.2
0.500 8.95 10.2 1.25 14.0
0.750 16.2 18.2 2.0 12.3
0.983 20.5 22.1 1.6 7.8
1.17 25 27.1 2.1 8.4
2.00 24.2 24.3 0.1 0.4
4.00 18.3 20 1.7 9.3
6.00 19 19.5 0.5 2.6
8.00 18.9 17.5 −1.4 −7.4
12.0 15.6 16.2 0.6 3.8
24.0 13.3 13.2 −0.1 −0.8
48.0 9.46 10 0.54 5.7
72.0 6.56 6.46 −0.10 −1.5
96.0 4.46 4.76 0.3 6.7
120 3.55 3.69 0.14 3.9
168 2.11 2.29 0.18 8.5
192 1.55 1.76 0.21 13.5
216 1.22 1.39 0.17 13.9