Table 1.

Acquisition and reconstruction parameters for EID- and PCD-CT scans performed with metal implants within the image field of view. * indicates the tube current of PCD-CT was adjusted to match the CTDIvol values of the clinical EID-CT scan for each patient. ** indicates the PCD reconstruction kernel was matched to the corresponding kernel for the EID system on which a specific patient was scanned. Thus, if the EID CT used iterative reconstruction (IR) in our clinical protocol (e.g., Flash scanner), we used IR for the PCD-CT images; if the EID CT used filtered back projection (FPB) in our clinical protocol (e.g., Sens64 scanner), we used FBP for the PCD-CT images.

Body
Part
Scanner
Type
kV Quality
Reference
Collimation Kernels Slice
Thickness
Cervical Spine Flash 140 280 mAs 128 × 0.6 mm I31(2), I70(2) 2.0 mm
Sens64 140 350 mAs 64 × 0.6 mm B31, B70 2.0 mm
PCD-CT 140Sn * 48 × 0.25 mm ** 2.0 mm

Lumbar Spine Flash 120 260 mAs 128 × 0.6 mm I31(2), I70(2) 2.0 mm
Sens64 120 350 mAs 64 × 0.6 mm B31, B70 2.0 mm
PCD-CT 140Sn * 48 × 0.25 mm ** 2.0 mm

Shoulder Edge 140 250 mAs 128 × 0.6 mm I40(2), I70(2) 1.0 mm
PCD-CT 140Sn * 48 × 0.25 mm I40(2), I70(2) 1.0 mm

Wrist Edge 140 200 mAs 128 × 0.6 mm I40(1), I70 (1) 0.6 mm
Edge 140 250 mAs 16 × 0.3 mm U40, U70 0.6 mm
PCD-CT 140Sn * 48 × 0.25 mm ** 0.6 mm

Elbow Edge 140 250 mAs 16 × 0.6 mm U40, U70 1.0 mm
PCD-CT 140Sn * 48 × 0.25 mm U40, U70 1.0 mm

Ankle Edge 140 250 mAs 16 × 0.6 mm U40, U70 1.0 mm
PCD-CT 140Sn * 48 × 0.25 mm U40, U70 1.0 mm