Box 1, steps 4, 6–9 |
Liquid is not spraying from the analytical column |
Column is not fully open |
Examine the column under a microscope; if it is not fully open, etch the tip by dipping it in HF for 1–2 min (see Box 1 for details. HF is extremely dangerous; follow the precautions outlined in Box 1). The column should also be examined for any obstructions (dust, particles) that may prevent it from packing |
|
|
Open end of the column is clogged |
The end of the column can become damaged or clogged when inserting it through the ferrule, leading to the high-pressure bomb. After insertion, cut ~0.2 cm off the end of the column using a ceramic scoring wafer |
Box 1, step 15 |
Column is packing too quickly with 3.5-|im packing material |
Slurry of 3.5-μm packing material is too concentrated |
Make a more dilute solution of packing material by increasing the amount of ACN |
Box 1, step 17 |
Column is not packing |
|
Turn helium off, and then back on, using the valve connected to the pressure bomb. This will sometimes cause the column to begin packing again |
|
|
Open end of the column is clogged |
The end of the column can become clogged with packing material. Cut ~0.2 cm off the end of the column using a ceramic scoring wafer |
|
|
Slurry of 1.7-μm packing material is too concentrated |
Ensure that all of the packing material is able to go into solution when stirred; if not, increase the amount of ACN in the slurry (note that ACN will evaporate over time) |
|
|
Open end of the column does not reach the packing material slurry |
Ensure that the end of the column is fully submerged in the packing material slurry |
|
Gaps in packing material |
Poor flow of packing material; blockage |
Increase the pressure of the helium |
|
|
|
Connect to UPLC and flow the 100% mobile phase B through the column at a flow rate of 0.3 μl/min for 15 min. Discard the column if the gap does not fill |
Reversed-phase Chromatography (Equipment Setup) |
High back-pressure during LC-MS/MS separation |
Analytical column may be clogged |
Replace the column |
|
|
High-pressure union is clogged |
Over time, the high-pressure union can become blocked with packing material or fused silica. Remove the union. When held up to the light, you should be able to see through it. If the union is blocked, wash it with ethanol to remove the debris. If necessary, replace the union |
|
|
Analytical column is attached too tightly at the union |
If the analytical column is attached too tightly, the end can become crushed and block the union. Remove the column and trim the end with a scoring wafer. To avoid this problem, after threading the column through the ferrule, trim the end with a ceramic scoring wafer. The column should protrude ~1 mm out of the ferrule. Tighten the analytical column as much as possible, without crushing the tubing. Once the analytical column is connected, gently pull on it. If it comes loose, use slightly more force when you are tightening the union |
|
|
25-μm line is clogged |
Verify by removing the analytical column; if there is not a drastic decrease in pressure, replace the 25-μm line |
|
High back-pressure during sample loading (>8,500 psi) |
Impurities in the sample |
Ensure that the sample was properly desalted before analysis |
Chromatography |
Very low back-pressure |
Column not properly attached |
Verify whether the liquid is flowing from the tip; if not, make sure that the 25-μm line and the analytical column are properly connected at the union. If no liquid flow is observed from the 25-μm line, check its connection to the LC system |
|
Broad peaks (>1 min) |
Column was packed >1 cm with 3.5-μm packing material |
Discard the column (see step 11 from Box 1) |
|
|
Gaps in packing material in column |
Flow 100% solvent A through the column at 300–350 μl/min (with column heater). Inspect the column; if there are no more gaps and the packed column is of suitable length, trim the unpacked end from the column using a ceramic scoring wafer. If gaps persist, discard the column |
|
|
Sample overloading |
Reduce the amount of material injected on the column |
|
Late-eluting sample |
Column degradation |
Inject 100–300 fmol of tryptic BSA peptides on the column; if peaks have shifted significantly, discard the column. Similarly, if total ion chromatogram (TIC) and base peak intensities are significantly lower than usual and the instrument is working as expected, discard the column |
|
|
Trapping sample for too long |
Decrease the trapping time according to the instructions in the ‘Reversed-phase chromatography’ section (Equipment Setup) |
|
|
Column is not packed all the way to the end |
Inspect the column using a light source. If it is not packed to the end, remove the unpacked portion using a ceramic scoring wafer |
|
|
25-μm line too long |
If possible, trim this line to <50 cm |
|
|
Pumps are not delivering the correct amount of solvent |
Measure the flow rate of pumps A and B; if they are not delivering the correct amount of solvent, the gradient may need to be adjusted to accommodate the actual flow rates. Contact the LC vendor for maintenance help |
|
Run-to-run variation |
Column is not fully equilibrated |
Increase the time of the equilibration period at the end of the LC gradient |
|
|
Pumps are not delivering the correct amount of solvent |
Measure the flow rate of pumps A and B; if they are not delivering the correct amount of solvent, the gradient may need to be adjusted to accommodate the actual flow rates. Contact the LC vendor for maintenance help |
|
|
Column degradation |
Inspect the chromatogram for wide peaks. This can also be tested by injecting 100–300 fmol of tryptic BSA peptides on the column. If peaks are wider than ~40 s, discard the column |
8 |
Amidation reaction |
Urea degradation during reduction (Step 7) |
Perform the reduction step (Step 7) at a lower temperature (i.e., at ambient temperature) |
24 |
Fewer identifications than expected; data-base search gives more semi-tryptic peptides than tryptic peptides |
Active endogenous proteases |
Collect yeast at an earlier time point. If this is not an option, try digesting with a higher enzyme: protein ratio for a shorter period of time (for example, 1:5 ratio for 4 h) (Steps 11,12) |
24 |
Large number (>50%) missed cleavages |
|
Increase the enzyme to protein ratio to 1:10 (Steps 11,12) |
14 |
Protein precipitates out of solution on acidification with TFA |
|
Use a less concentrated solution of TFA; add TFA slowly, by checking the pH after each addition to avoid overacidification |
24 |
Fewer identifications than expected |
Instrument is out of calibration |
Follow the recommended calibration schedule provided in the calibration console. If MS mass error is >6 p.p.m., recalibrate |
|
Fewer identifications than expected; decrease in TIC intensity over time |
Quadrupole may require maintenance |
Over time, the quadrupole and other parts of the instrument may require cleaning or maintenance. Contact your service engineer for instructions |