Table 5.

Differences Between Audiometric Pure Tone Thresholds and EC OR Thresholds at Different Time Points (TSO—1–3 days After Surgery; FF—5 Weeks After Surgery).

250 Hz
500 Hz
1000 Hz
Mean ± standard deviation in dB, correlation
Preoperative audiogram vs. preinsertion OR ΔECOR-Audio 47.0 ± 23.1 r= .32 25.2 ± 24.2 r= .10 8.0 ± 10.1 r= .76*
EC OR before vs. after insertion ΔECORpost-preins 5.5 ± 8.6 r= .62 2.1 ± 14.7 r= .42 6.0 ± 7.0 r= .88***
Preoperative audiogram vs. postinsertion EC OR ΔECORpost-Audiopre 52.5 ± 26.7 r= −.10 27.3 ± 20.4 r= .25 14.0 ± 9.4 r= .60
Postoperative audiogram (TSO) vs. postinsertion EC OR ΔECORpost-AudioTSO 20.5 ± 32.0 r= −.26 −7.2 ± 22.6 r= .29 −15.5 ± 12.8 r= .40
Postoperative audiogram (FF) vs. postinsertion. EC OR ΔECORpost-AudioFF 22.0 ± 39.5 r= −.36 −6.7 ± 30.1 r= −.16 −16.0 ± 16.3 r= .05

Note. OR = ongoing response; EC = extracochlear; FF = first fit; TSO = test switch-on.

*

p < .05 (significant). ***p < .001 (highly significant).