Table 2.

Alerts sent to investigational sites by the central monitoring unit in IN-TIME: definitions and occurrencesa

Alert type Device/HMSC: alert setting CMU activity in a patient with HMSC alert CMU: condition for SOP compatible event Number of alerts (number of patients)
Atrial tachyarrhythmia 109 (65)
 First onset of AF for >30 s in patients with no prior documented AF Device: IEGM for atrial monitoring episode Check IEGM for atrial monitoring episode in patients with no prior documented AF IEGM shows true AF 31 (31)
HMSC: sends alert
 Long atrial arrhythmia episode (>6 h) with high-ventricular rate (>120 b.p.m.) HMSC: alert for atrial burden >6 h with ventricular rate >120 b.p.m. Check IEGM Ventricular rate >120 b.p.m. None
 Daily AF burden >50% on 7 consecutive days after a period of 4 weeks with burden <25% for patients with known AF HMSC: alert for atrial burden >6 h Check trend graph Atrial burden ≥50% on 7 consecutive days 20 (16)
 Atrial arrhythmia in the VT zone HMSC: VT zone detected Check IEGM Suspicious IEGM 1 (1)
 First onset of AF for >30 s in patients with prior documented AF Device: IEGM for atrial monitoring episode Check IEGM for atrial monitoring episode Not predefined 22 (22)
HMSC: alert for atrial monitoring episode
 Atrial episode HMSC: alert for atrial monitoring episode Check IEGM for atrial monitoring episode Not predefined 22 (14)
 Atrial burden >50% for more or less than 1 week HMSC: alert for atrial burden ≥6 h Check trend graph Not predefined 13 (8)
Ventricular tachyarrhythmia or ICD shock 56 (42)
 ≥3 VT2/VF episodes over 48 h HMSC: alerts for VT2 and VF detection Check IEGM for appropriateness and count true VT/VF episodes in 48 h ≥3 episodes within 48 h 7 (6)
 First ICD shock in spontaneous episode HMSC: alert for VT and VF detections Check if the episode was induced or spontaneous Spontaneous episodes with shock 22 (22)
 First occurrence of slow VT (<150 b.p.m.) HMSC: alert for VT1 (monitoring zone) Check IEGM if true VT True VT <150 b.p.m. 8 (8)
 VT/VF detection with suspicious IEGM (e.g. sensing issue or inappropriate detection) HMSC: alert for VT and VF detections Check IEGM Suspicious IEGM 2 (2)
 <3 VT/VF episodes over 48 h HMSC: alert for VT and VF detections Check IEGM Not predefined 18 (14)
 Ineffective ICD shock HMSC: ineffective shock Check IEGM Not predefined 2 (2)
 VT and low CRT pacing HMSC: alert for VT and low CRT Check IEGM Not predefined 1 (1)
Low percentage of CRT pacing 91 (35)
 CRT pacing <80% over 48 h HMSC: alert for CRT pacing <80% Check CRT pacing on the day before CRT pacing <80% on two consecutive days 81 (30)
 CRT pacing <80% combined with IEGM HMSC: alert for CRT pacing <80% Check IEGM Suspicious IEGM 3 (3)
 CRT pacing <80% over 24 h HMSC: alert for CRT pacing <80% Check trend graph Not predefined 8 (7)
 CRT pacing <80% and ventricular or atrial arrhythmia HMSC: alert for CRT pacing <80% Check IEGM Not predefined 2 (2)
Ventricular extrasystoles 54 (46)
 Visible upward trend in VES/h over 7 days Not programmable Check trend graph weekly Visible trend 39 (34)
 VES >110 per hour for more than 10 days Not programmable Check trend graph Not predefined 15 (15)
Physiological trends 2 (2)
 Visible downward trend in patient activity over 7 days Not programmable Check rend graph weekly Visible trend 1 (1)
 Rising mean heart rate Not programmable Check trend graph Not predefined 1 (1)
Lead related measurements and IEGMs 98 (58)
 Suspicious IEGMb HMSC: alert for any new IEGM Check IEGM Suspicious IEGM 25 (18)
 RA amplitude HMSC: programmable Check trend graph <0.5 mV 13 (9)
 RV amplitude HMSC: programmable Check trend graph <2.0 mV 22 (18)
 RV pacing threshold HMSC: programmable Check trend graph Safety margin <1 V 11 (10)
 RV impedance HMSC: programmable Check trend graph <250 or >1500 Ohm 1 (1)
 LV amplitude HMSC: programmable Check trend graph <2.0 mV None
 LV impedance HMSC: programmable Check trend graph <250 or >1500 Ohm 2 (2)
 LV pacing threshold HMSC: programmable Check trend graph Safety margin <1 V 12 (8)
 Shock impedance HMSC: programmable Check trend graph <30 or >125 Ohm 13 ( 5)
Gaps in data transmission 818 (241)
 Missing HM messages for 3 days HMSC: alert for missing messages for 3 days Check if the patient is known to be on holidays or in hospital If the patient is not known to be absent 818 (241)
Technical alerts None
 Elective replacement indicator HMSC: programmable None
 Technical alert HMSC: programmable None

Sums of numbers of alerts do not match completely because few alerts are contained in more than one category, and three alerts contained no description of the content.

CMU, central monitoring unit; ICD, implantable cardioverter-defibrillator; HM, Home Monitoring; SOP, standard operating procedure defined in the study protocol; VT, ventricular tachycardia; AF, atrial fibrillation; CRT, cardiac resynchronization therapy; IEGM, intracardiac electrogram; LV, left ventricular; RA, right atrial; RV, right ventricular; VES, ventricular extrasystoles; VF, ventricular fibrillation; VT1/VT2, slow/fast ventricular tachycardia.

a

During the randomized period in the RM group.

b

T-wave oversensing, far-field atrial sensing of ventricular activity, or other suspected sensing problem.