Skip to main content
. 2019 Mar 28;57(4):e01336-18. doi: 10.1128/JCM.01336-18

TABLE 1.

Limits of detection of the plasma separation card for cobas HIVb

Concn (copies/ml) No. of valid samplesa No. of reactive samples Detection rate (%)
1,971.1 62 62 100.0
1,925.3 63 63 100.0
1,358.1 62 62 100.0
985.5 63 62 98.4
962.6 63 62 98.4
679 63 59 93.7
657 63 59 93.7
641.8 63 57 90.5
452.7 63 52 82.5
328.5 63 47 74.6
320.9 63 46 73.0
226.3 62 49 79.0
164.3 63 36 57.1
160.4 63 35 55.6
113.2 63 38 60.3
0 63 0 0.0
0 63 0 0.0
0 63 0 0.0
a

Some sample sizes were <63 due to an insufficient number of spots. Data for limit of detection analysis for CAP/CTM are not shown.

b

Limit of detection by probit analysis (95% hit rate), 790.2 copies/ml. 95% confidence interval (CI), 658.9 to 1,003.6 copies/ml. Lowest panel member with ≥95% reactive rate, 962.6 copies/ml 98.4% (62/63); 95% CI = 91.5 to 100.0%; lower one-sided 95% CI = 92.7%.