Table 2.

Electrical characteristics and patient follow up (FU).

Patient number Implant time (mins) Fluoroscopy (sec) Number of deployments R wave (mV) Threshold (V at 0.24 ms) Impedance (Ω) Pacing mode Symptoms at follow up
1 25 460 2 15.4 0.38 680 Hyst 40/70 No
2 70 300 2 7.4 3.5 800 VVI 50 No
3 60 243 1 4.4 0.5 680 VVI 40 No
4 46 475 1 6.1 0.38 820 Hyst 40/50 No
5 n/a n/a 2 10.1 0.5 760 VVI 50 No FU
6 34 234 1 5.7 0.25 670 Back up No
7 40 420 1 7 0.63 620 Hyst 30/60 No
8 50 511 1 18.7 0.5 1500 VVI 40 No
9 48 619 2 7.5 0.5 550 VVI 60 No
10 40 269 1 9.2 0.25 810 Hyst 40/70 No FU
11 40 608 2 15.4 0.38 750 VVI 40 Yes
12 20 195 1 7.7 0.63 1430 Hyst 40/70 No
13 49 608 1 9.9 0.38 780 Hyst 40/80 No
14 42 752 2 12.9 0.63 620 Hyst 50/70 No
15 30 534 1 10.9 0.88 730 Hyst 30/80 No
16 26 512 1 6.2 0.5 670 Hyst 40/80 No
17 20 190 1 10.7 0.25 1150 Hyst 40/70 No
18 41 230 2 16.3 1.88 750 VVI 35 No
19 24 319 1 6.1 0.88 890 Hyst 40/70 No
20 40 278 2 5.3 0.375 790 Hyst 40/70 No
21 44 344 2 22.7 0.5 710 VVI 40 No
22 49 500 1 9.9 0.375 780 Hyst 40/80 Yes
23 93 1467 1 9.5 0.5 520 Hyst 40/80 No
24 40 218 1 8.9 1.63 610 Hyst 40/80 No
25 n/a 470 3 4.3 0.5 1100 VVI 40 No
26 n/a 108 1 7.7 0.38 700 VVI 40 No
27 n/a 282 1 20 0.63 760 VVI 40 No
28 n/a 282 2 13.6 0.38 1270 VVI 40 No
29 n/a n/a 1 4.8 0.5 660 VVI 40 No
30 n/a 948 3 7.8 0.75 750 VVI 40 No
31 n/a 384 1 13.4 0.5 710 VVI 40 No
32 35 39 1 9.8 9.8 740 VVI 40 No