Skip to main content
. Author manuscript; available in PMC: 2019 Apr 3.
Published in final edited form as: Diagn Interv Imaging. 2017 Aug 24;99(2):99–104. doi: 10.1016/j.diii.2017.07.010

Table 3.

Average flow rate (mL/min) of water and 30% diluted ultrasound gel (viscous fluid) with particulate matter (TVP) with control and different connections. The number of occluded trial and unpredictability of which catheters and connection devices would become occluded prevented summarizing the data with descriptive statistics and analyzing with inferential statistics. The data are presented to illustrate the inconsistencies, which may be indicative the occlusive nature and unpredictability of particulate matter/debris.

Catheter size (Fr) Control Luer-lok (2.67mm) Low pressure (2.33mm) Standard (2.00mm) High pressure (1.33mm)
Water and TVP
8.5 occluded in 3/10 trials occluded in 5/10 trials occluded in 3/10 trials occluded in 8/10 trials occluded in 2/10 trials
10.2 occluded in 1/10 trials occluded in 8/10 trials 55±41 69±35 occluded 10/10 trials
12.0 23 ±30 occluded in 6/10 trials occluded in 7/10 trials 33±54 occluded in 9/10 trials
14.0 6±2 occluded in 7/10 trials 41±38 occluded in 6/10 trials occluded in 3/10 trials
Viscous fluid and TVP
8.5 occluded in 4/10 trials occluded in 2/10 trials occluded in 4/10 trials occluded in 4/10 trials occluded in 5/10 trials
10.2 occluded in 1/10 trials 27±25 occluded in 1/10 trials occluded in 1/10 trials occluded 1/10 trials
12.0 occluded in 2/10 trials occluded in 1/10 trials occluded in 3/10 trials occluded in 5/10 trials occluded in 6/10 trials s
14.0 9±9 occluded in 1/10 trials 10±8 occluded in 2/10 trials occluded in 7/10 trials
14.0 6±2 occluded in 7/10 trials 41±38 occluded in 6/10 trials occluded in 3/10 trials

Note. The number of occluded trial and unpredictability of which catheters and connection devices would become occluded prevented summarizing the data with descriptive statistics and analyzing with inferential statistics. The data are presented to illustrate the inconsistencies, which may be indicative the occlusive nature and unpredictability of particulate matter/debris.