Table 3.
Average flow rate (mL/min) of water and 30% diluted ultrasound gel (viscous fluid) with particulate matter (TVP) with control and different connections. The number of occluded trial and unpredictability of which catheters and connection devices would become occluded prevented summarizing the data with descriptive statistics and analyzing with inferential statistics. The data are presented to illustrate the inconsistencies, which may be indicative the occlusive nature and unpredictability of particulate matter/debris.
| Catheter size (Fr) | Control | Luer-lok (2.67mm) | Low pressure (2.33mm) | Standard (2.00mm) | High pressure (1.33mm) |
|---|---|---|---|---|---|
| Water and TVP | |||||
| 8.5 | occluded in 3/10 trials | occluded in 5/10 trials | occluded in 3/10 trials | occluded in 8/10 trials | occluded in 2/10 trials |
| 10.2 | occluded in 1/10 trials | occluded in 8/10 trials | 55±41 | 69±35 | occluded 10/10 trials |
| 12.0 | 23 ±30 | occluded in 6/10 trials | occluded in 7/10 trials | 33±54 | occluded in 9/10 trials |
| 14.0 | 6±2 | occluded in 7/10 trials | 41±38 | occluded in 6/10 trials | occluded in 3/10 trials |
| Viscous fluid and TVP | |||||
| 8.5 | occluded in 4/10 trials | occluded in 2/10 trials | occluded in 4/10 trials | occluded in 4/10 trials | occluded in 5/10 trials |
| 10.2 | occluded in 1/10 trials | 27±25 | occluded in 1/10 trials | occluded in 1/10 trials | occluded 1/10 trials |
| 12.0 | occluded in 2/10 trials | occluded in 1/10 trials | occluded in 3/10 trials | occluded in 5/10 trials | occluded in 6/10 trials s |
| 14.0 | 9±9 | occluded in 1/10 trials | 10±8 | occluded in 2/10 trials | occluded in 7/10 trials |
| 14.0 | 6±2 | occluded in 7/10 trials | 41±38 | occluded in 6/10 trials | occluded in 3/10 trials |
Note. The number of occluded trial and unpredictability of which catheters and connection devices would become occluded prevented summarizing the data with descriptive statistics and analyzing with inferential statistics. The data are presented to illustrate the inconsistencies, which may be indicative the occlusive nature and unpredictability of particulate matter/debris.