Skip to main content
. 2018 Jun 25;9(2):173–178. doi: 10.1177/2192568218784298

Table 3.

Quantification of a Typical Manufacturer’s Reprocessing Guideline.

Process Step Minimum Time (Minutes) Type Allocation For 164 Devices and 1 Person (Guideline) Real-Time Observation
Inspection 1 0.15 Visual Per device 24.6 N/A
Precleaning 2 1 Running water Per device 164 N/A
3 2 Cleaner and brush Per device 328 N/A
4 0.1 Water All 0.1 N/A
5 1 Water jet Per device 164 N/A
6 15 Ultrasonic cleaner All 15 N/A
7 2 DI water jet Per device 328 N/A
Inspection 8 0.15 Visual Per device 24.6 N/A
Mechanical washer 9 2 Prewash All 2 4
10 2 Wash I All 2 12
11 5 Wash II All 5
12 2 Rinse All 2 5
13 40 Dry All 40 20
Thermal disinfection 14 2.5 At 93°C All 2.5 8
Inspection 15 0.15 Visual Per device 24.6 N/A
Packaging for sterilization 16 N/A Wraps All 3 3
Sterilization 17 4 Steam All 4 4
Drying 18 20 Up to 60 minutes All 20 20
Total steps 18 Total time 19 hours 1 hour 16 minutes