Beam matching criteria currently used for Elekta linacs with Agility heads in this study. Each parameter is compared to a reference. Profile points are taken to be within 80% of the field size

Modality Parameter Tolerance
Photons PDD20,10 0.5%
PDD points 0.5%
Profile points 1.0%
Wedge profile points 2.0%
Electrons R50 0.5 mm
Profile points 1.0%