. 2019 Mar 18;20(4):99–105. doi: 10.1002/acm2.12567

Table 2.

Matched beam parameters for linacs 1 to 9. Profiles assessed for a 30 × 30 cm2 field at 10 cm depth in the in‐plane (IP) and cross‐plane (XP) directions. PDDs are taken for a 10 × 10 cm2 field at either 100 or 90 cm SSD for beam quality matching

Linac Flatness (%) Symmetry (%) Quality (90 SSD) Quality (100 SSD) Output
# Model IP XP IP XP PDD20,10 TPR20,10 PDD20,10 TPR20,10 (cGy/MU)
1 VersaHD 103.8 104.4 100.3 100.8 0.574 0.666 0.996
2 Synergy 103.9 104.0 100.1 100.1 0.588 0.684 1.003
3 VersaHD 104.8 104.4 101.1 100.6 0.574 0.669 0.998
4 VersaHD 103.8 103.7 100.0 100.0 0.573 0.665 0.996
5 Synergy 104.3 104.5 100.7 100.8 0.585 0.682 1.005
6 VersaHD 105.1 104.9 100.9 100.7 0.576 0.669 0.589 0.687 0.995
7 Synergy 104.0 103.8 100.8 100.7 0.588 0.685 1.000
8 Synergy 104.1 103.9 100.7 100.1 0.587 0.684 1.005
9 VersaHD 104.0 103.9 100.3 100.4 0.572 0.665 1.000
Standard deviation 0.5 0.4 0.4 0.3 0.002 0.003 0.007 0.009 0.004
Baseline 104.8 104.4 100.0 100.0 0.574 0.669 0.588 0.685 1.000

TPR: tissue phantom ratio.