Published in final edited form as: J Am Soc Echocardiogr. 2019 Apr;32(4):537–544.e3. doi: 10.1016/j.echo.2019.01.015

Table 2.

Intraclass correlation coefficients (ICCs) and limits of agreement demonstrating repeatability of measurements of right ventricular (RV) function. In all groups, the number of subjects tested was 20. (TAPSE: tricuspid annular plane systolic excursion, FAC: fractional area change, S’: peak tricuspid annular systolic tissue velocity.)

Intra-reader Reliability: Transplants ICC (95% Confidence Interval) Limits of Agreement Mean Difference (95% Confidence Interval)
TAPSE by M-mode 0.97 (0.94–1.00) −0.09 – 0.14 0.025 (−0.003 – 0.052)
TAPSE by 2D 0.79 (0.63–0.96) −0.18 – 0.34 0.078 (−0.018 – 0.138)
TAPSE by M-mode and 2D 0.95 (0.90–0.99) −0.74 – 0.75 0.003 (−0.171 – 0.177)
FAC 0.85 (0.72–0.97) −0.08 – 0.07 −0.005 (−0.022 – 0.012)
S’ 0.99 (0.97–1.00) −0.48 – 0.59 0.057 (−0.068 – 0.181)
Intra-reader Reliability: Controls
TAPSE by M-mode 0.95 (0.91–0.99) −0.11 – 0.35 0.123 (0.069 – 0.176)
TAPSE by 2D 0.93 (0.86–0.99) −0.37 – 0.38 0.004 (−0.083 – 0.091)
TAPSE by M-mode and 2D 0.92 (0.85–0.99) −1.17 – 0.97 −0.100 (−0.350 – 0.150)
FAC 0.77 (0.58–0.95) −0.06 – 0.12 0.033 (0.012 – 0.055)
S’ 0.93 (0.88–0.99) −0.95 – 1.28 0.165 (−0.095 – 0.424)
Inter-reader Reliability: Transplants
TAPSE by M-mode 0.88 (0.79–0.98) −0.27 – 0.35 −0.031 (−0.086 – 0.024)
TAPSE by 2D 0.88 (0.74–0.98) −0.31 – 0.20 −0.056 (−0.115 – 0.004)
FAC 0.76 (0.57–0.95) −0.06 – 0.05 −0.002 (−0.014 – 0.011)
S’ 0.89 (0.80–0.98) −1.68 – 0.81 −0.438 (−0.730 - −0.146)
Inter-reader Reliability: Controls
TAPSE by M-mode 0.97 (0.94–1.00) −0.30 – 0.13 −0.086 (−0.135 - −0.036)
TAPSE by 2D 0.84 (0.72–0.97) −0.29 – 0.58 0.145 (0.043 – 0.247)
FAC 0.56 (0.26–0.87) −0.08 – 0.09 0.006 (−0.014 – 0.026)
S’ 0.92 (0.85–0.99) −1.19 – 1.18 −0.006 (−0.283 – 0.271)