Table 2.
The results of diagnostic test by visual assessment of the CP OCT images.
Polarization | Inter-rater reliability rate | Sensitivity, % | Specificity, % | Diagnostic accuracy, % |
---|---|---|---|---|
Co- | 0.83/0.83 | 89–93 | 67–73 | 83–84 |
Cross- | 0.86/0.86 | 80–87 | 75–89 | 82–83 |
Co- and cross- simultaneously | 0.92/0.91 | 82–85 | 92–94 | 87–88 |