Skip to main content
. 2019 Apr 9;23:100358. doi: 10.1016/j.ijcha.2019.100358

Table 2.

Procedural results.

Characteristic Asian Non-Asian p-Value
Successful implant 99.1%
(105/106)
97.9%
(93/95)
0.60
LAA seal
 Complete seal or Jet size < 5 mm 100.0% (105/105) 100.0% (93/93) NS
 Jet size ≥ 5 mm 0.0%
(0/105)
0.0%
(0/93)
LAA diameter
 N 106 92
 Mean ± SD 23.4 ± 4.1 21.2 ± 3.2 <0.0001
 Median 23.00 21.50
Last device size used (mm)
 Mean ± SD 27.4 ± 3.4 25.3 ± 3.2 <0.0001
 Median 27 24
Compression of last device size used (mm)
 Mean ± SD 17 ± 7% 17 ± 6% 0.93
 Median 16% 17%

p-Value = sample t-test across regions. LAA = left atrial appendage, mm = millimetres, NS = not significant.

Wilcoxon Rank-Sum test used to compare medians.