Table 1.

Solid angle parameters for the three beam sizes used in the measurement. The first column gives the diameter of the upstream/ downstream apertures in units of mm; the second column is the number of Dy images that were acquired; the third column is the solid angle Ω¯AG of the alpha–gamma alpha detector; and the last column is the resulting efficiency correction.

Collimation Number of images Ω¯AG ϵ correction (Ω¯AG/ΩAG(0,0))
15/7 2  0.007 0964 ± 0.000 0003 0.992 049 ± 0.000 04
15/8 3  0.007 0788 ± 0.000 0003 0.989 587 ± 0.000 05
15/10 5  0.007 0517 ± 0.000 0004 0.985 798 ± 0.000 05