Table 7.
Component | Comment | Type | ui%2 | |
---|---|---|---|---|
Activity ratio | Standard uncertainty of combined activity ratio measurements from this study | A | 0.48 | |
LTAC 229Th activity | Standard uncertainty of the M solution massic activity | B | 0.28 | |
229Th molality | Combined standard uncertainty of the measured 229Th molality | B | 0.07 | |
Dilution factor | Standard uncertainty of the dilution factor for production of SRM 4328C from the M solution | B | 0.03 | |
Decay interval | Relative uncertainty associated with the interval between the M solution reference date and molality measurements |
B | 0.00 | |
Relative combined standard uncertainty (uc) | 0.56% | |||
229Th half-life (a) | Expanded uncertainty3 (a) | |||
7825 | 87 (Relative: 1.1%) |
The uncertainty estimate for the half-life was calculated in accordance with Taylor and Kuyatt, [20] and GUM protocols [21] using the GUM Workbench software [22]. ui% represents the relative standard uncertainty for the listed components. The expanded uncertainty (U = k × uc) for the half-life value is the product of the combined standard uncertainty (uc) and a coverage factor (k), with k = 2 to achieve an approximately 95% level of confidence