Summary of the average uncertainty estimates for the LJO analyzer during 2015. Each component of the total measurement uncertainty is listed, where is the total mean annual uncertainty in the air measurements collected during 2015 and calculated as described by Eqs. (5)–(7), is the average annual uncertainty in the enhancement, Uh2o is the uncertainty due to the treatment of water vapor, is the mean uncertainty derived from the target tank measurements, is the mean analyzer precision, is the mean calibration baseline uncertainty, and is the extrapolation uncertainty, or the uncertainty due to the single-point calibration strategy, which was estimated using a mean ε for nine analyzers (see text and Supplement).