Figure 8:

Figure 8:

Representative LenStar measurement of a mouse eye showing that the calculation of VCD results in a negative number if the LenStar internal analysis (i) is used for this species: VCD = 1.84–0.117–0.26–1.72–0.046 = −0.303 mm.