. 2019 Jan 31;4(2):233–236. doi: 10.1016/j.adro.2019.01.008

Table 1.

In vivo peripheral dose measurements with optically stimulated luminescence dosimeters for chin, pelvis, and breast areas

Measurement point Measured dose per fraction (cGy) Total dose for 20 fractions (cGy) TPS predicted total dose (cGy) Difference between TPS and measurement (%)
Chin 1 2.04 40.8 - -
Chin 2 2.03 40.6 - -
Chin 3 2.05 41 - -
Pelvis 1 0.81 16.2 - -
Pelvis 2 0.85 17 - -
Pelvis 3 0.86 17.2 - -
Breast (average of 2 highest measurements) 109.13 2182.6 2242.5 −2.70%

Abbreviation: TPS = treatment planning system.

Comparisons with the treatment planning system predictions are shown where available.