Skip to main content
. Author manuscript; available in PMC: 2019 Apr 13.
Published in final edited form as: Abdom Radiol (NY). 2016 May;41(5):844–853. doi: 10.1007/s00261-015-0589-3

Table 1.

PIRADS v2 recommended parameters for the DCE MRI sequence

DCE sequence parameters 2D or 3D T1 gradient echo (3D preferred)
TR/TE <100 ms/<5 s
Slice thickness 3 mm, no gaps
Field of view Entire prostate gland & seminal vesicles
In plane dimension ≤2 mm×≤2 mm
Temporal resolution ≤15 s (<7 s preferred)
Total observation time ≥2 min
Contrast dose 0.1 mmol/kg standard GBCA or equivalent high relaxivity GBCA
Contrast injection rate 2–3 cc/s with continuous image acquisition

Adapted from ACR PIRADS v2 [23]