Skip to main content
. 2019 Feb 7;6(8):1802128. doi: 10.1002/advs.201802128

Figure 3.

Figure 3

a) Measured I–V curves with different temperatures and pressures. The temperature of one Peltier element was constantly kept at 22 °C while varying the other one. b) Voltage axis intercept of I–V curves (thermal voltage output) as a function of time, under different humidity and temperatures. c) The peak Seebeck coefficient as a function of mechanical pressure, measured under RH = 90% and RH = 60%. d) The resistance of PTH sensor as a function of humidity, measured under P = 100 Pa and P = 300 Pa. The error bars are standard deviations obtained from at least five independent measurements.