Table 1.
Real-time PCR (+) | Real-time PCR (−) | No. of clinical samples | Sensitivity, 95% CI (%) | Specificity, 95% CI (%) | Overall agreement, 95% CI (%) | |
---|---|---|---|---|---|---|
On-chip LAMP (+) | 53 | 2 | 100 | 100 · 0, [93 · 3–100 · 0] |
95 · 7, [85 · 5–99 · 5] |
98 · 0, [93 · 0–99 · 8] |
On-chip LAMP (−) | 0 | 45 |
Assuming real-time PCR results (not a gold standard method) are correct.