Table 2. Summary of anti-EBOV GP IgG ELISA qualification results.
Parameter | Result |
---|---|
Negative Sample OD | 95% Prediction Interval of 0.000–0.337 |
Limit of Detection | 31.74 ELISA units/mL |
Lower Limit of Quantitation | 55.34 ELISA units/mL |
Upper Limit of Quantitation | 2,511.89 ELISA units/mL |
Intermediate Precision | 14.8% CV |
Repeatability | 8.2% CV |
Dilutional Linearity (Accuracy) | Slope of -0.99 (90% CI of -1.03, -0.94) |
rGP-Coated Plate Stability | 7 days at 2–8°C |
Serum Stability | 21 days at 2–8°C 24 hours at room temperature 7 freeze/thaw cycles |
rGP Stability | 3 days at 2–8°C 8 freeze/thaw cycles (6 recommended) |