Skip to main content
. 2019 Mar 13;9(3):429. doi: 10.3390/nano9030429

Table 3.

Determination of 4-NP in water samples (n = 4).

Type of Water a Added (μM) Found by This Method b (μM) Recovery (%) Found by HPLC b (μM) Relative Error (%)
River 0.5 0.512 (±0.015) 102.4 0.483 (±0.012) 6.00
Lake 1.0 0.976 (±0.018) 97.6 1.024 (±0.022) −4.69
Waste 2.0 2.071 (±0.033) 103.6 1.962 (±0.042) 5.56
Tap 0.1 0.102 (±0.002) 102.0 0.106 (±0.003) −3.77

a All water samples were collected from Hengyang, China. b average ± confidence interval, the confidence level is 95%.